小型光學氧微傳感器配件專為各種定制應用而設計。微型探頭的尺寸< 50 μm 至 140 μm,外徑范圍為 140 μm 至 900 μm。微型傳感器未安裝在任何額外的外殼中。裸露的玻璃纖維可以安裝到您自己的外殼、鋼管和微型呼吸計室等上。小外徑甚至允許插入植入式Venflon管中。此外,微型傳感器可以部署在密封容器中,以直接測量氧氣含量。IMP-PSt1 的測量范圍為 0 – 50 % 氧氣,檢測限為 0.05 % O2.
· 高空間分辨率 (< 50 μm)
· 高時間分辨率 (t90< 3 秒)
· 集成到定制外殼和設置中
· 無需耗氧
· 頂空和液體測量
小型光學氧微傳感器配件
TECHNICAL
Specifications | Gaseous & Dissolved O2 | Dissolved O2 |
*after two-point calibration as described in the manual | ||
Measurement range | 0 – 50 % O2 0 – 500 hPa | 0 – 22.5 mg/L 0 – 700 µmol/L |
Limit of detection | 0.05 % oxygen | 20 ppb |
Resolution | ± 0.01 % O2 at 0.21 % O2 ± 1 hPa at 207 hPa | ± 0.005 mg/L at 0.09 mg/L ± 1.3 µmol/L at 283 µmol/L |
Accuracy* | ± 0.4 % O2 at 20.9 % O2 | |
Drift | < 0.1 % O2 within 30 days (sampling interval of 1 min.) | |
Measurement temperature range | from 0 to + 50 °C | |
Response time TS** (t90) | < 1 sec. (gas) | < 2 sec. (liquid) |
Response time TF*** (t90) | < 15 sec. (gas) | < 30 sec. (liquid) |
Properties | ||
Compatibility | Aqueous solutions, ethanol, methanol | |
No cross-sensitivity | pH 1 – 14 | |
Cross-sensitivity | Organic solvents, such as acetone, toluene, chloroform or methylene chloride | |
Sterilization procedure | Steam sterilization | |
Cleaning procedure | 3 % H2O2 | |
Calibration | Two-point calibration with oxygen-free environment (nitrogen, sodium sulfite) and air-saturated environment | |
Storage stability | 60 months provided the sensor material is stored in the dark at room temperature |